5
Technical Information
ION Pump Controller Description

Remote I/O Connector Pinout

Figure 6
Remote I/O Pinout Rear Panel D-Shell Female 25P
Installation according to international rules.
All the input/output remote signals to/from the controller must be
connected at mating connector. With the provided mating connector
2
make the connection with AWG 24 (0.25 mm
) or smaller wire to the
pins indicated in the figure to obtain the desired capability. It is a 15-
pins D type connector; the available signals are detailed in the table,
the following paragraphs describe the signal characteristics and use.
The controller is shipped with the mating (male) connector with the 3
jumpers on the HV 1,2,3 On/Off (External Interlock) inputs in order
to allow to switch the HV On and Off by the unit front panel.
In case there is the need to use the Remote Interlock function or to
switch On and Off the HV on the 3 pumps by Remote, remove the 3
jumpers and connect the appropriate connections.
Ion Pumps Control Unit (No Battery) User Manual / 87-900-116-01 (E)
85/101