5
Technical Information
Procedure to Connect the Serial and I/O Ports to an External Cable
HV Channel Start/Protect Mode Setting
"C0ns"

Clear Fault (Reset)

"F0n"

Fault Table

Err. nb
[01]
[02]
[04]
[08]
[10]
[20]
[40]
[80]
94/101
where :
C0 = HV Channel Serial Command for Start/Protect Setting
n = HV Channel Number 1, 2 or 3
s = 0 for START mode, 1 for PROTECT mode
where:
F0 = HV Channel Serial Command for Fault reset
N = HV Channel number 1,2,3
Description
Cable Interlock
Remote Interlock
HV Overvoltage
ADC Offset high
Output short circuit
Loc/Ser switch moved
Overtemperature
Overcurrent in Protect
Ion Pumps Control Unit (No Battery) User Manual / 87-900-116-01 (E)
Front Panel Led
1 blink every 2 sec.
2 blink every 2 sec.
3 blink every 2 sec.
4 blink every 2 sec.
5 blink every 2 sec.
6 blink every 2 sec.
7 blink every 2 sec.
8 blink every 2 sec.