4
Installation Procedure

Operating Modes

Operating Modes

Power Line Supply

LOCAL-REMOTE I/O Operation

The 3 HV switches on the front panel and the 3 HV On/Off inputs on the DB25
NOTE
"Remote" connector on the rear panel are in series.
The HV to the 3 pumps can be switched ON by the 3 HV (HV1, HV2,
HV3) switches on the front panel (in this case the DB25 plug with the
internal jumpers provided with the unit must be connected and the 3
HV Cable Interlocks must be closed) or by the closure of the 3 HV
On/Off inputs on the DB25 "Remote" connector on the rear panel (in
this case the 3 HV switches on the front panel must be in the ON
position).
When the HV is ON, the corresponding HV LED on the front panel,
will be ON.
If the connection between the IPCU and the pump is done by means of an HV
NOTE
Cable with Safety Interlock, the cable must be connected to the pump F/T and
the small banana connector on the cable controller side must be inserted in the
corresponding "HV Cable Interlock" plug on the unit rear panel in order to be
able to switch on the HV.
If the connection is done with a single pole HV Cable without the
Safety Interlock, ensure that the jumper provided on the unit rear
panel is inserted in the "HV Cable Interlock" plug.
76/101
Ion Pumps Control Unit (No Battery) User Manual / 87-900-116-01 (E)