Controller Report Mode

"HVn status fault start/protect current voltage"
HV
n
' '
status
' '
fault
' '
start/protect [0, 1] 0=Start(default), 1=Protect 1 character
' '
Current
' '
voltage

Controller Commands Mode

HV Channel Status Setting
"A0ns"
Ion Pumps Control Unit (No Battery) User Manual / 87-900-116-01 (E)
Procedure to Connect the Serial and I/O Ports to an External Cable
[HV]
[1, 2, 3]
[0x20]
[OFF__, FAULT, ON___]
[0x20]
[00, .., FF]
[0x20]
[0x20]
[10 ... 99999nA, 100 ... 50000uA] 7 characters
[0x20]
[0 ... 5999V]
[0x0D]
where :
A0
= HV Channel Serial Command for Status Set
n
= HV Channel Number 1, 2 or 3
s
= 0 for OFF, 1 for ON
Technical Information
2 characters 'HV'
1 character
1 character SPACE
5 characters
1 character SPACE
2 characters
1 character SPACE
1 character SPACE
1 character SPACE
6 characters 'CR '
1 character CR
Total 30 characters
93/101