4
Installation Procedure

Installation

Installation

WARNING!
The controller is equipped with a 3-wire power cord without plug. Install the
proper plug foreseen by the country where the controller will be used.
Note that the Yellow/Green wire of the cable is the Ground connection of the
controller. Use this power cord and the installed plug in conjunction with a
properly grounded power socket to avoid electrical shock. Verify that the
ground connections are properly connected. High voltage developed in the
controller can cause severe injury or death. Before servicing the unit,
disconnect the power cable.
The controller can be used as a bench unit or as a rack module, but it must be
CAUTION!
positioned so that free air can flow through the holes of the top and the side
cover. If the controller is used as a rack module, it MUST be inserted in a 4 unit
height rack adapter chassis to avoid the controller to fall inside the rack. The
Controller front panel is not designed to support the unit weight. Do not install
or use the controller in an environment exposed to atmospheric agents (rain,
snow, ice), dust, aggressive gases, or in explosive environments or those with a
high fire risk.
If the controller is installed on a rack, remove the four feet and position it with
NOTE
at least 30 mm (1.2 inches) of clearance on top and bottom.
WARNING!
The controller must be installed in a way that allows an easy interruption of
the line voltage (disconnection of the line plug or interruption on the rack
general line switch).
68/101
Ion Pumps Control Unit (No Battery) User Manual / 87-900-116-01 (E)