With Battery supply:
Output Voltage
Output Current (total)
Output Power
Operating Temperature
Storage Temperature
Compliance to Norms
Line Cable
Line fuse
High Voltage Connection
Battery Connection
Installation according to international rules, only of the following HV cables Part No.
may be used:
HV Cables P/N
929-0705
929-0707
929-0708
929-0709
Ion Pumps Control Unit (No Battery) User Manual / 87-900-116-01 (E)
Technical Information
ION Pump Controller Description
-3000 Vdc +20, - 5 %
0.3 mA
1 W
0 °C to 40 °C
-20 °C to 70 °C
Radio Interference: EN55011 Class A
Group 1
Harmonics:
Flickers:
ESD:
Burst:
Radiated RF Immunity: EN61000/4/3
Surge:
Conducted noise imm.: EN61000/4/6
Line supply interruptions:EN61000/4/11
Safety:
3 m long cable (3 wires, 3 m long), no plug
Two T5A
3 High Voltage Fischer connectors
3 pin connector (for Lead Acid Battery 12
Vdc 24 Ah)
HV bakeable cable, radiation resistant,
4 meter
HV bakeable cable, radiation resistant,
7 meter
HV bakeable cable, radiation resistant,
10 meter
HV bakeable cable, radiation resistant,
20 meter
5
EN61000/3/2
EN61000/3/3
EN61000/4/2
EN61000/4/4
EN61000/4/5
EN61010-1
83/101