If the control result is not inside the acceptable range, here are some things you can do to
solve the problem:
Troubleshooting Checks
1. Check if the test strips or control
solutions were expired.
2. Check if you wiped the tip of the
control solution bottle before and after
use.
3. Check if the caps on the test strip
container and the control solution
bottle were always closed tightly.
90
Action
Discard the test strips or control solution
if either is past the expiry date. If the
control solution was opened more than
three months ago, discard it.
Wipe the tip of the bottle with a tissue.
Repeat the control test with a new test
strip and a fresh drop of control solution.
Replace the test strips or control
solutions if you think either was
uncapped for some time.