Operating conditions

To ensure that your CoaguChek XS System functions properly, please
observe the following guidelines:
Only use the meter at a room temperature between 15 °C and 32 °C.
Only use the meter at a relative humidity between 10% and 85%.
When testing, place the meter on a level, vibration-free surface or
hold it so it is roughly horizontal.
If the meter is to remain unused for a longer period of time, keep it in
the carry-case supplied.
Use the meter only up to a maximum altitude of 4300 metres.
Electromagnetic interference
Strong electromagnetic fields may interfere with the
proper operation of the meter. Do not use the meter near
strong electromagnetic fields.