Camille Bauer Digem f 96x48CK Mode D'emploi page 139

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

3.5
External Control Inputs
!
Attention!
Terminals 8, 9 10 and 11 are electrically connected to the measurement input. Isolation
of external circuit elements must be executed in a fashion suitable for the prevailing
measurement input potential.
Instrument Test (Test)
When terminals 10 and 11 are bridged,
the entire display is darkened.
Attention: This connection causes reset-
ting of the microprocessor. All stored min.
and max. values, as well as automatic tare
values, are lost.
After this connection is interrupted, a seg-
ment test is conducted for about 1 second.
Thereafter, the instrument returns to normal operation.
Display Storage (Hold)
When terminals 9 and 11 are bridged in the normal display mode, the currently displayed value is
stored. This has no influence on the measuring sequence.
In the counter mode, this causes setting of the counter to zero. In the tare mode the tare zero-value
is set.
Program Protection (Lock)
When terminals 8 and 11 are bridged, the limit values selected before adjustment are protected and
the programming of important parameters is disabled. If an attempt is made to change the protected
limit values selected prior to adjustment, the message "Loc" appears in the display.
3.6
Serial Interface (only for DIGEM f 96 x 48 CK)
Connector Pin Assignment
RJ-11
Socket
6 5 4 3
at DIGEM
A connector cable is included with measuring instruments with serial interface.
GOSSEN-METRAWATT
1 2 3 4 5 6
RJ-11
Plug
3 4 5 6
8 9 10 11 12
Cable, 1.5m
5
4
3
2
9
8
7
6
9 Pole Sub-D Port for PC
14 15
Test
Hold
Lock
1
7

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Digem f 96x48ek

Table des Matières