6.2.1
Checking the Concentration Setting
Required auxiliaries:
In the following example which shows how to determine the nominal concentration, the
quantity settings given in millimeters refer to a tapped water quantity of exactly 8 liters.
Formulas for setting the concentration:
Explanantion:
K
= Concentration (%)
VD
= Volume of disinfectant (ml)
VW
= Volume of water (ml)
Example:
Disinfectant quantity to be set according to the above formula:
To check the concentration setting, proceed as follows:
Unscrew the flow control unit from the mixing chamber and seal the mixing chamber
opening with the supplied sealing plug.
Insert the flow control unit into the provided measuring jar and place a bucket below the
swivel drain outlet.
Press the ON/OFF (I/0) switch and tap approx. 2-3 liters of water.
Empty the bucket and the measuring jar.
Insert the flow control unit back into the measuring jar and place the bucket below the
swivel drain outlet again.
Start the unit by pressing the ON/OFF (I/0) switch.
Tap exactly 8 liters of water. Switch off the unit (press the ON/OFF (I/0) switch).
Read the disinfectant quantity collected in the measuring jar.
Calculate the concentration according to the above formula. If it deviates from the
desired concentration (plus buffer), adjust the setting as described
in chapter
Remove sealing plug from mixing chamber opening and screw flow control device in
again.
6.2.2
Checking the Safety Equipment
A check of the safety equipment obligatory includes the following test procedures:
functional test of the level suction pipe
functional test of the flow control device
6.2.3
Functional Check of the Level Suction Pipe
For the functional test of the suction pipe "empty" signal device, proceed as follows:
Pull suction pipe out of product supply container.
After approx. 3 seconds, unit should change to fault mode (red fault indicator flashes).
417101158_DG1.docx
Bucket with a minimum capacity of 10 l to collect the water
Measuring jars to collect the disinfectant:
up to 100 ml for concentrations of 1 % max.
up to 300 ml for concentrations of 1 % to 3 %
Desired concentration = 0.5 %, plus 10 % buffer
K x VW
VD =
100 − K
6.2
„Setting the Concentration".
VD x 100
K =
Formula:
VW + VD
0,55 x 8.000
=
= ,
100 − 0,55
40
K x VW
VD =
100 − K
= 0.55 %
Rev. 07-01.16