CAUTION
• H old the spray can upright.
• B e sure to hold the handpiece firmly to prevent slipping out of hand by the spray pressure when
lubricating.
• S upply lubricant until it comes out of the handpiece head.
NOTICE
• N SK recommends the use of "Spray Mist Absorber" (Y900084) to prevent oil mist expelling out of
the handpiece head.
NSK automatic handpiece cleaning and lubrication system
When using NSK automatic handpiece cleaning and lubrication system, refer to the system instructions.
WARNING
• I f blood infiltrates inside a handpiece, an automatic handpiece cleaning and lubrication system may
not totally clean the internal handpiece components, which may lead to internal blood coagulation.
Coagulated blood inside the handpiece may cause handpiece failure and overheating, causing burn
injuries. To avoid the risk, NSK recommends to immediately lubricate those handpieces with PANA
SPRAY Plus / PANA SPRAY.
20-5 Sterilization
Sterilize the product by autoclave sterilization. After each patient sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20min. at 121°C, or 15min. at 132°C, or 3min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.
CAUTION
• B e sure to clean and lubricate the handpiece prior to autoclaving. If blood remains on the internal
surface it can become clotted and cause product failure.
• D o not autoclave the product with other instruments even when it is in a pouch. This is to prevent
possible discoloration and damage to the product from chemical residue on other instruments.
• D o not heat or cool the product too quickly. Rapid change in temperature could cause damage to the
product.
• T o avoid product failure, do not use a sterilizer whose temperature exceeds 138°C during use
including the dry cycle. The chamber temperature of some type of sterilizers may exceed 138°C.
Contact the sterilizer manufacturer for detailed information.
• K eep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight.
The air should be free from dust, salt and sulphur.
• D o not touch the product immediately after autoclaving as it will be very hot and must remain in a
sterile condition.
• A utoclave sterilization is recommended for the product. The validity of other sterilization methods is
not confirmed.
NOTICE
• N SK recommends Class B sterilizers as stated in EN 13060.
21. Periodical Maintenance Checks
Perform periodical maintenance checks every three months, referring to the check sheet below. If any
abnormalities are found, contact your Authorized NSK Dealer.
Points to check
Rotation
Rotate the handpiece and check for abnormalities such as abnormal rotation,
vibration, noise, and overheating.
Details
15