Table 2 Setup menu
Variable
Description
SAMPLE
Shows a second menu, Sample Text menu. Under this menu,
LABEL
set the following items:
• SAMPLE LABEL "SAMPLE NAME"—The user-input name for
samples. The sample labels limit is 10 characters. For each
sample label, there is an auto-counter extension that increments
up to 500. This extension is reset to one every time a new
sample label is programmed, re-entered or exceeds a count of
500.
• REMARKS—User-input remarks for any additional sample
labeling, as necessary. These optional remarks become part of
the record of any completed samples. Records retrieved from
the sample buffer can have their remarks reviewed and re-
edited as necessary. The sample remarks limit is 10 characters
on any of 3 lines. Any existing remarks show on printed sample
reports.
MODE
Sampling method (BOTTLE or ONLINE or FILTER)
If BOTTLE mode is selected or re-entered, a second screen is
shown and it allows the following variable to be set.
• PURGE VOL—Volume of fluid that flows through the
instrument to purge any contamination before samples are
taken (15–30 ml).
If ONLINE mode is selected or re-entered, a second screen is
shown and it allows the following variable to be set.
• NUM SAMPLES—Numbers of samples that run during the
sampling period (1–500).
• HOLD TIME—Time delay from the end of a sample to the start
of the next sample. If set for 00:00, a minimum 5-second hold
time is set (hh:mm).
• PURGE VOL—The volume of fluid that flows through the
hydraulic hose and unit to purge any contamination before
samples are taken. The longer the hydraulic hose, the greater
the purge volume must be (15–999 ml).
STD
Controls the standard format to show the sample results
(ISO(C), ISO, NAS(C), NAS, NAV(C), NAV, SAE(C), SAE, MIL,
P/10 ml or P/1 ml).
Table 2 Setup menu (continued)
Variable
Description
VOL
Volume of fluid that runs 3 times during a sample
(Volume/RUN). The BOTTLE mode sampling is limited (5, 10 or
20 ml per RUN).
FLOW
The flow rate is fixed at 50 mL/min.
PRINT
Enables or disables the printer so that it does not automatically
print at the end of a sample. The SETUP option prints the
current instrument setup (ENBL, DSBL or SETUP).
CONTRAST
Use the navigation keys to control the display contrast.
Table 3 Screen: PG 2
Variable Description
LANG
Shows the language used in the operation and control of the
instrument (English, French, German, Span and Ital).
TEMP
Shows the temperature units (°C or °F).
PRESS
Shows the pressure units (BAR or PSI).
DISPL
Controls the brightness of the background: light background (Normal)
or dark background (Reversed) (NORM or REV).
BKLT
Controls the back light on the LCD display. If AUTO is selected, the
back light turns off after 5 minutes of inactivity (ON, OFF or AUTO).
BEEP
Controls audio feedback of the beeper when a key is pushed (KEY or
DSBL).
BSAVE
Battery save feature. If enabled, the instrument automatically turns off
after 15 minutes of no activity (Idle state) ((ENBL (Enabled) or DSBL
(Disabled)).