Specifying Cuff Settings; Selecting A Measurement Mode - Philips SureSigns VS1 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

If for some reason the monitor does not detect arterial occlusion, it will inflate to the value specified in the
Initial Inflation Pressure setting, which is described below.
Note—
If a patient is moving, shaking, or agitated during a blood pressure measurement, the Smart
Inflation feature might interpret the movement as a pulse. If this occurs, the cuff could potentially inflate to
a pressure higher than 180 mmHg (23.9 kPa). If your patient cannot remain still, Philips recommends that
you disable the Smart Inflation feature. To disable Smart Inflation, press the System soft key, and then press
the Cuff Option soft key. Use the OK button to move the cursor to the Smart Inflation choice and uncheck
the X. When Smart Inflation is disabled, the monitor will not inflate any higher than the value specified in
the Initial Inflation Pressure setting.

Specifying Cuff Settings

Use the Cuff Settings screen to specify the overall parameters of the NIBP cuff measurements. The cuff
pressure is shown in the mean arterial pressure (MAP) window during the measurement cycle.
MEASUREMENT MODE: ADULT/PEDI
INIT INFLATION PRESSURE:
[X] BP UPON ALARM
[X] SMART CLOCK
[X] SMART INFLATION
To display the Cuff Settings screen and change the cuff setting values:
Step
Action
1
From the Line List display, press the System soft key. The System Settings screen appears.
2
From the System Settings screen, press the Cuff Option soft key.
3
Use the OK button to select the cuff setting you want to change. The cuff settings are described below.
4
Use the Up or Down button to change the value of a selected setting.
To save your changes, press either the System Screen soft key or the Save → Main soft key. If you press
5
the System Screen soft key, the settings are saved and the System screen appears. If you press the Save →
Main soft key, the settings are saved and the Line List display appears.

Selecting a Measurement Mode

The Measurement Mode has two settings — Adult/Pediatric and Neonatal. After you select a mode and
save it, the Cuff Mode Indicator light on the front of the monitor displays the currently selected mode:
Adult/Pediatric
Neonatal
180 mmHg
SYSTEM
SCREEN
Specifying Cuff Settings
11:06
SAVE
->MAIN
Monitoring Blood Pressure
5-3

Publicité

Table des Matières
loading

Table des Matières