System Assembly

Warning
Prior to first use the equipment must be processed as per the included Anspach G1 High Speed System
Cleaning and Sterilization instructions.
Note: When following these instructions hold handpiece and attachments with distal end pointing away from
user.
Console, Handpiece and Foot Control Assembly
1. Place console on a flat sturdy surface.
Warning: Console must be outside sterile field.
2. Plug hospital grade AC power cord into AC power inlet on back of console. Plug opposite end of power cord
into standard, Hospital Grade grounded wall outlet.
Caution: Connect to supply mains with protective earth only. Position equipment in such a manner that it
provides access to the rear of the system in order to disconnect from mains
3. Insert foot control connector into foot control connector port on front of console with red dot on connector
facing up.
Warning: Foot control must be outside sterile field.
Caution: Do not push foot control connector into console connector port when out of alignment.
4. Insert handpiece connector into handpiece connector port on front of console with red dot on connector
facing up.
Warning: Handpiece must remain inside sterile field with exception of handpiece connector which is
connected to console.
Caution: Do not push handpiece connector into console connector port when out of alignment.
5. Activate console by depressing power switch located on front of console. Display will illuminate.
6. Refer to Operating Instructions below.
Warning: Handpiece is fully functional at this time.
Caution: Do not use handpiece without an attachment and dissection tool properly locked in place. Damage to
the burr lock mechanism could occur.
Attachment Assembly
1. Refer to the Anspach G1 Attachment instructions accompanying the specific device.
Dissection Tools
1. Refer to the Anspach G1 Dissection Tools instructions accompanying the specific device.
Optional Irrigation Tubing Set-Up (SC3002 only)
1. Remove sterile irrigation tube from sterile package.
2. Slide irrigation tubing onto proximal end of irrigation attachment clip.
3. Attach irrigation attachment clip to handpiece.
4. Attach irrigation attachment clip to attachment.
5. Route irrigation tube end into non-sterile area.
6. Open twist-lock for irrigation pump in direction of arrow shown.
7. Insert irrigation tube into pump in accordance with marking and close twist lock.
8. Insert IV pole into bracket on back of console.
9. Hang irrigation bag onto IV Pole (maximum 1 L bag).
10. Remove protective cap from cannula and connect cannula to irrigation bag. When doing so, make sure
connection nipple of cannula is not touched by non-sterile persons while securing irrigation.
11. Secure tubing to handpiece hose using Irrigation Tube Hose Clips.
12. To activate, depress "Irrigation" button on front of console. Light on front of console will be on to indicate
irrigation is active.
13. To control flow of irrigation, rotate knob located adjacent to irrigation pump.
14. Irrigation can also be activated from certain optional Foot Controls (see Step 4 of Foot Control Operation).