System Shutdown

1. There are no special procedures for system shutdown. Depress power switch on front of console to shut down
system for approximately 2-3 seconds.
2. To disconnect power cord from mains, grasp ridged portion of connector between thumb and forefinger of one
hand. Gently pull connector away from console. Connector should remove easily. If not, ensure only ridged portion
of connector is pulled. Do not pull on smooth part of connector, as connector will not disengage from console.

Additional Information

To isolate the device from the mains supply, remove the power cord. All specifications are subject to change.
The eG1 System operates continuously without a duty cycle under normal operating conditions. When a A-1 or H-1
alert is indicated on the Console display the following Duty Cycle must be employed:
1 minute ON / 1 minute OFF for unlimited number of cycles for ambient temperatures up to 29°C (85° F).
After completing the procedure perform the Inspection and Maintenance steps provided. If the alert does not clear,
return the system for repair to DePuy Synthes Power Tools.
Potential Equalization Terminal
The purpose of additional potential equalization is to equalize potentials between different metal parts that can be
touched simultaneously, or to reduce differences of potential which can occur during operation between the bodies of
medical electrical devices and conductive parts of other objects. The eG1 System Console incorporates a dedicated
potential equalization terminal that is identified by the symbol and is designed as a means of establishing potential
equalization.
The potential equalization terminal is specially shaped so as to prevent the risk of accidental disconnection when used
as intended and allows the lead to be removed without the need for tools. The terminal is located at the back of the
eG1 System console and is accessible to the operator during normal use.
The potential equalization terminal on the eG1 System Console cannot be used as a protective earth connection.
The eG1 System Console provides an IEC320 inlet with a dedicated protective earth blade. The power supply cord
provided with the eG1 System Console does not provide a potential equalization conductor.
Prior to using the eG1 System, attach the potential equalization wire to the potential equalization terminal at the back
of the eG1 System Console.

End of Life

Return all equipment for proper disposal.
Recommended Manufacturer Inspection Interval
It is recommended that the equipment be returned to DePuy Synthes Power Tools at a minimum every 12 months to
perform a full product inspection.