Console Alerts, Codes, and Displays
Result
Turn OFF Alert Sounds
Turn On Alert Sounds
Foot Pedal Connection
Foot Pedal Connection
Handpiece Connection
Console Status
Console Status
Console Status
Software Version Revision Forward and Reverse buttons are pressed
Stall Condition
System Fault
Handpiece Fault
Console Fault
Temperature Alert
Airflow Alert
Action
Press speed increase, speed decrease, and
forward buttons simultaneously
Press speed increase, speed decrease, and
forward buttons simultaneously
Foot Pedal is unplugged, or if the console is
turned on without a foot pedal
Foot pedal is plugged in , or if the console is
turned on with a foot pedal connected
Handpiece is connected, or if the console is
turned on with a hand piece connected
Console is ON, but not ready to run, because
either the Handpiece or Foot Pedal, or both are
not connected
Console ready to run
Console that has a connected Handpiece
Running at full speed
simultaneously.
NOTE: Example Only
Handpiece does not start when the foot pedal is
depressed.
• Check Attachment Lock
• Release Foot Pedal and restart
Handpiece does not start when the foot pedal is
depressed
• Release Foot Pedal and restart
The Handpiece must be detached from the
console.
If the alarm clears after the hand piece is
removed, the hand piece is damaged.
The Console is damaged and must be returned
for repair.
Handpiece temperature exceeded.
Handpiece cooling air flow is blocked.
• Must use Duty Cycle.
Display
9