8-7 Sterilization
Sterilize the product by autoclave sterilization. After each patient sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20min. at 121°C, or 15min. at 132°C, or 3min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.
CAUTION
• Be sure to clean and lubricate the handpiece prior to autoclaving. If blood remains on the internal surface it can become clotted and cause
product failure.
• Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to the
product from chemical residue on other instruments.
• Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
• To avoid product failure, do not use a sterilizer whose temperature exceeds 138°C during use including the dry cycle. The chamber
temperature of some type of sterilizers may exceed 138°C. Contact the sterilizer manufacturer for detailed information.