Télécharger Imprimer la page

Rs232-1 And Rs232-2; Glp Data - OHAUS Adventurer Pro AV53 Guide De L'utilisateur

Masquer les pouces Voir aussi pour Adventurer Pro AV53:

Publicité

Les langues disponibles

Les langues disponibles

Adventurer Pro
Content
All of these features can be set On or Off.
Numeric data only, Header, Gross, Net, Tare,
Reference, Result, GLP. (See sample printout
3.10)
Layout
Determines the format of data output to a printer
or computer. If Line Format is set to Multi, a
multi-line printout is generated. If it is set to
Single, a single line printout is generated. If 4
LF is set to yes, 4 line feeds are appended to
the printout. If Form feed is set to Yes, a form
feed is appended to the printout. This is useful
for printing to page printers.
List
When Yes is selected, a printout of balance
settings is generated.
End Print-1
Press the Yes button to advance to the next
menu, Print-2 if COM2 is installed or RS232-
1 if COM2 is not installed. Pressing No returns
to Output menu item.

3.7.7 RS232-1 and RS232-2

The RS232-1 submenu is used to set
communication parameters for an external
printer or computer.
RS232-2 contains an identical menu and is
available when the optional COM2 port is
installed.
RS232-1
Power
On/Off
Baud
600..2400...19200
Parity
7 Even, 7 Odd, 7 No
Par, 8 No Parity
Handshake Off, XONXOFF
Hardware
End RS1
Power
This menu item allows setting the Power On or
Off for COM1. It is not available for COM2.
When the balance is operated from the AC
Adapter, this menu item is hidden and the
setting is On.
When the balance is operated from batteries,
the menu item is available and the default
setting is Off. To enable COM1, Power must
be turned On.
Baud
Baud rates of 600, 1200, 2400, 4800, 9600
and 19,200 are available.
Parity
Parity settings of 7 even, 7 Odd, 7 No Parity
and 8 No Parity are available.
Handshake
Settings of Off, XONXOFF and (for RS232-1
only) Hardware are available.
End RS1 (RS2)
Pressing the Yes button will advance to the RS232-
2 (or GLP Data) submenu. Pressing No returns
to the Power (or Baud) menu item.

3.7.8 GLP Data

GLP DATA
User ID
Set...
Proj ID
Set...
Time
Type-12hr, 24hr,
Set..., Adj-60, +60
Date
Type mdy,...dym
Set...
End GLP Data
User ID
Used to enter a User ID up to 10 characters.
EN-23

Publicité

Dépannage

loading