Span Calibration

Refer to 2.5 for span calibration.

Linearity Calibration

WARNING
Only required if the linearity error exceeds
tolerance in specification table.
Select LINEARITY from the CAL sub-menu.
Press the Yes button and follow screen
instructions. Use the weight values shown
on the primary display.
After calibration, remove weight from pan.

3.7.2 Setup

SETUP
Legal Trade On/Off
Auto Tare
On/Off
Auto Off
Off, 1, 2, 5 min.
Back Light
On/Off
End Setup
Legal Trade
When set On, the balance operates in compliance with Weights and Measures regulations.
Auto Tare
When set On, the balance will automatically tare the first item placed on the pan.
Auto Off
When set to 1, 2 or 5 minutes, the balance will turn off in the selected time interval if there has
been no activity. Set to OFF for continuous operation.
Back Light
Used to set the display back light on or off.
End Setup
Press the Yes button, to advance to the Readout submenu. Press the No button to return to the
LEGAL TRADE menu item.