8� Troubleshooting Guide
Error
Problem
message
Autotest in progress, device not ready.
Ambient temperature is not within
measurement range.
Forehead temperature above 108 °F (42.2 °C)
or surface temperature above 176 °F (80 °C).
Forehead temperature below 93.2 °F (34 °C)
or surface temperature below 32 °F (0 °C).
Batteries low or depleted.
Solution
Wait for forehead symbol to stop
flashing.
Place the thermometer in an ambient
temperature between 59 °F and 104 °F
(15 °C - 40 °C) for at least 30 minutes.
Only use thermometer within
measurement range.
Only use thermometer within
measurement range.
Replace batteries.