Tecno-gaz Europa B evo Mode D'emploi page 87

L'autoclave
Masquer les pouces Voir aussi pour Europa B evo:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

14.2 Connection of demineralizer system
Turn off the autoclave by pressing the General switch (MAINS) located on the right side of
the autoclave. (PICT.2 pos..17.)
The operator after having set the autoclave can install the demineralizations system
following the indications in the manual attached to the demineralizer apparatus.
Here are indicated the specific connection to the autoclave of the water supply hose and
the connection of the electric plug.
1. Turn off the autoclave if it is on;
2. Close the faucet located above the demineralization system;
3. Install the demineralizer as indicated in its manual;
4. Twist the male screw of the connection-hose carrier with Teflon or other component
which guarantees the sealing of water;
5. Screw the connection on the Demineralizer tap PICT. 32 pos. .2.;
6. Insert the hose exit of demineralizer to the connection-hose carrier already screwed to
the autoclave;
7. Insert the plug of demineralizer to the socket PICT.32 pos. .7. on the back of the
autoclave
8. Open the faucet located above the demineralization system;
9. Control that there is no water loss;
10. Turn on the autoclave;
11. Run one or more sterilization cycles to control the functioning of the connection and
check above all for leaking.
At the end of each day always close the faucet located above
demineralization system to prevent flooding
Connect of demineralizer only to preprogrammed autoclave .
If the unit at the beginning of the cycle shows e250 then you have to fill the
clean water tankuntil the maximun is reached.
For the connection of demineralization system to the autoclave refer to
NOTE
indications in the manual.
14.3 Change resins - filter
Autoclave display indicates the state of finishing of the resin with the message "REPLACE
THE RESINS" or "REPLACE THE FILTER".
Substitute the resin cartridge as indicated in the deminaralizer manual.
After substitution, press button
until the "beep", to reset the autoclave.
B-PUMP
ENGLISH
the
39

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Europa b evo 24

Table des Matières