Example 1 Data Logger Mode; Example 2 Data Logger Mode - Giorgio Bormac 8+ Série Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

PC
The recorded data is directly sent to the PC through USB connection and processed using the
appropriate software DataLink+.
P5.2 Data logging type
Enter this menu to select the acquisition mode data.
Manual
: The data is acquired or printed only when the user presses the
Seconds / Minutes / Hours
range. Press
key to start and end the registration.
P5.3 Print format
Enter this menu to select which information to print in the header in a data logger printer mode.
For 8+ Series:
• Simple: progressive registration number / date and hour / value, UM and temperature.
• Complete: device model / report last cal / progressive registration number / date and hour /
value, UM and temperature.
For 80+ Series:
• Simple: progressive registration number /
ID sample / date and hour / value, UM and
temperature.
• Complete: model and sn of the device /
date last cal / IDs / progressive registration
number / date and hour / value, UM and
temperature.
• GLP: model and sn of the device/ report last
cal / IDs / progressive registration number /
date
and
hour
temperature.

• Example 1 Data Logger mode

Example for automatic pH recording on internal memory every 2 minutes.
Enter in Data Logger setup mode.
In P5.1 select Memory.
In P5.2 select Minutes, with directional keys set "2" and confirm by pressing
key to return in measurement mode and activate MEAS pH screen. In the upper band you
Press
can find icons
- saving in internal memory - and
Press
to start the registration,
progress. The number beside
Press again
key to end the recording.
Note: automatic recording is suspended when you change the measurement parameter or enter the setup
menu.
Note2: recording and printing in simultaneous pH/Cond screen occur for both the values.

• Example 2 Data Logger mode

Example for manual printing of Cond value with date of last calibration.
Enter in Data Logger setup mode.
In P5.1 select Printer.
In P5.2 select Man.
In P5.3 select Complete.
UM Serie 880 EN rev.2 30.07.2020
: By using the arrow keys set an automatic data acquisition frequency
/
value,
UM
and
icon starts blinking, meaning that the memorisation is in
icon indicates how many data has been saved.
30
key.
key.
- automatic data acquisition -.
EN