P1.8 Restore factory default.
If the instrument does not properly work or incorrect calibration were performed press Yes with
return all pH parameters to the factory default settings.
IMPORTANT: the factory reset of the parameters does not delete stored data.

• Automatic pH calibration

Example for 3 points USA buffers calibration
In pH measurement mode press and hold for 3 sec "
In PC devices pH calibration can be accessed also from simultaneous pH/Cond screen then selecting pH.
Rinse the electrode with distilled water
Press
and immerse the electrode in the pH buffer 7.00 (as indicated by the becker on the display).
The first calibration point is always the neutral pH (7.00 for USA buffer, 6.86 for NIST and DIN buffers)
while the others are at the discretion of the user.
When the
icon appears on the display confirm the first calibration point pressing
The display flashes the value measured and then the pH 7.00 icon becker appears at the bottom left
7.00
indicating that the instrument is calibrated at the neutral point.
Extract the electrode, rinse with distilled water
Press
key to continue the calibration and immerse the electrode in the pH buffer 4.01.
In the becker all the values that the device can recognize flow.
When the value on the becker stabilizes on pH 4.01 and the
key. The display flashes the value measured and subsequently the pH 4.01
alongside the pH 7.00 becker, the device is calibrated in the acid range.
Extract the electrode, rinse with distilled water
Press
key to continue the calibration and immerse the electrode
in the pH buffer 10.01. In the becker all the values that the device is
able to recognize flow.
When value on the becker stabilizes on pH10.01 and the
appears confirm by pressing
The passage from an acid to a basic pH may take a few more seconds to achieve stability.
The display flashes the value measured and subsequently the pH 10.01 icon becker
alongside the pH 7.00 and pH 4.01 icon beckers, the device is calibrated also in alkaline range.
Although the device could accept two other calibration points, we stop and confirm this calibration
pressing
(8+ Series will automatically end after three points).
UM Serie 880 EN rev.2 30.07.2020
CAL" key to enter into calibration mode.
and gently dab with absorbent paper.
and gently dab with absorbent paper.
and gently dab with absorbent paper.
key.
16
key.
icon appears confirm by pressing
4.01
icon becker appears
icon
10.01
EN
key to
appears