icon reappears press
on the display and the becker icon appears with the identifier color and the buffer value at the bottom
U 4.65
left
.
Although the device could accept two other calibration points, we stop and confirm this calibration
(8+ Series will automatically end after three points).
pressing
The calibration report and graphic representation are displayed, press
measurement mode. At bottom left the buffers used for the last calibration are displayed, the
value is preceded by the letter "U" indicating that the value was entered manually.
Note: If you are working with manual temperature compensation (MTC), update the value before calibrating
the device.

• Performing pH measurement

Enter in the Setup menu for pH to check the calibration and if necessary, update the reading
parameters (see paragraph "pH parameters setup"), press
key to scroll through the different parameter screens until you find MEAS pH (see paragraph
Press
"Operation of the device").
The pH measurement can be displayed on three different screens:
• Digital
Connect the electrode to the BNC for pH/mV/ORP of device (yellow/blue).
If the user does not use an electrode with incorporated temperature probe or an external NTC 30KΩ
probe, it is advisable to manually update the temperature value (MTC).
Remove the electrode from its protection cap, rinse with distilled water and gently dab with absorbent
paper.
Check the presence and eliminate any air bubbles in the bulb of the membrane by vertically shaking
(as for the clinical thermometer). If it is present, open the side pad.
Immerse the electrode in the sample while maintaining slight stirring.
Consider truthful the measurement only when
interpretation it is possible to use "HOLD" function (P6.8) which allows to freeze the measurement as
soon as it achieves stability.
After the measurement, wash the electrode with distilled water and
store it in the appropriate storage solution. Never store sensors in
distilled water.
In the function "Display Complete P6.7", the graphic
representation of the electrode indicates the slope of the
current calibration.

• DHS Electrode

Electrodes equipped with DHS technology can save a
calibration curve inside their memory. The calibrated sensor
is automatically recognized by any DHS instrument and it
acquires the calibration.
Connect DHS electrode to the BNC and RCA of device.
UM Serie 880 EN rev.2 30.07.2020
key to confirm the second point, the actual measured value flashes
• Analogic
icon appears. To eliminate any errors due to user
18
or
to exit and return in
to return in measurement mode.