pH parameters

pH8+DHS, pH80+DHS, PC8+DHS, PC80+DHS
On this series of devices, you can use pH sensors with integrated temperature probe or connect two
different sensors. The pH electrode uses a BNC connector while the temperature probe uses a RCA/CINCH
connector. The instrument is also able to recognize DHS sensor, an innovative electrode capable of storing
the calibration data and then being immediately used on any enabled instrument.
• pH parameters setup
In measurement mode press "
Use
and
keys to scroll until "pH" menu and press
Use
and
keys to move and then select the program you want to edit.
The table below shows the structure of the setup menu for the pH parameter, for each program you can
find the options that the user can choose and the default value.
Program
P1.1
pH buffer selection
P1.2
Resolution pH
P1.3
Last calibration pH data
P1.4
Due calibration pH
P1.5
Set stability criteria
P1.6
Set pH alarm
P1.7
Temperature calibration
P1.8
Restore factory default
* function available only for 80+ Series
P1.1 pH buffer selection
Access this setup menu to choose the family buffer for calibrating the electrode.
The 80+ Series device allows the pH calibration to be performed from 1 to 5 points, while 8+Series can
perform calibrations from 1 to 3 points.
During the calibration press
calibration chart).
The instrument automatically recognizes 3 buffers family (USA, NIST and DIN*) in addition, the user
has the possibility to calibrate manually up to 5 points with customizable values (2 points for 8+ Series).
USA buffers: 1,68 - 4,01 - 7,00** - 10,01 - 12,45 (factory default)
NIST buffers: 1,68 - 4,00 - 6,86** - 9,18 - 12,46
1,68 - 4,01 - 6,86** - 9,18 - 12,45
DIN buffers:
** The neutral point is always required as the first point
In measurement mode at the bottom left of display a set of
becker indicates the tampons with which the last calibration was
made both automatically and manually. Inside the becker the
number represents the exact value of the buffer, and for a quick and intuitive understanding a color scale
has been inserted.
P1.2 Resolution pH
Access this menu to choose the resolution you want to get in measurement and calibration pH of the
parameter:
0.1
0.01 -factory default-
0.001 (available only for 80+ Series)
UM Serie 880 EN rev.2 30.07.2020
MENU" key to enter SETUP menu.
Description
USA – NIST – DIN* – USER
0.1 – 0.01 – 0.001*
View data – Print
NO – 1...99h – 1...99d
Tit – Med - High
NO – MIN - MAX
YES – NO
YES – NO
key to exit and save the calibrated points up to that moment (see
14
key to confirm.
Options
Factory default
View data
Becker color
Buffer pH value
Brown
Red
2.5 ~ 6.5
Green
6.5 ~ 7.5
Blue
7.5 ~ 11.5
Black
USA
0.01
NO
Med
NO
-
NO
< 2.5
> 11.5
EN