Mode Description - DeVilbiss Healthcare DV55 Bilevel Série Mode D'emploi

Masquer les pouces Voir aussi pour DV55 Bilevel Série:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN
Display Example
IPAP
10.0 cmH2O r 5.0
EPAP
5.0 cmH2O r 5.0
Delay Time
45 minutes
Delay Pressure
5.0 cmH2O
IPAP Rounding
0 (1234 ms)
EPAP Rounding
0 (1234 ms)
Insp. Trigger
5
Exp. Trigger
5
Backup Rate
15 BPM
I/E Ratio
1 : 1.0 (50%)
SE-DV56
14
aDVancED mEnu controlS
Mode
Description
This is a prescription setting which controls the
All
pressure during inhale, this also shows the
Bilevel
pressure delta (5.0 in example) between inhale and
Modes
exhale pressure settings
This is a prescription setting which controls the
All
pressure during exhale, this also shows the
Bilevel
pressure delta (5.0 in example) between inhale and
Modes
exhale pressure settings
Selectable delay time setting. Your provider has set
the delay time. If the unlocked symbol is shown,
you are able to adjust this setting. If the locked
All
symbol is shown, this setting can only be adjusted
by your provider.
This is a prescription setting which controls the
CPAP
operating pressure during the delay time.
Mode
This is a setting which controls the comfort
All
rounding when switching to IPAP pressure, settings
Bilevel
are 0 thru 5 also shows measured rise time of the
Modes
last breath in milliseconds.
This is a setting which controls the comfort
All
rounding when switching to EPAP pressure,
Bilevel
settings are 0 thru 5 also shows measured rise time
Modes
of the last breath in milliseconds.
This is a setting which controls the sensitivity of
Bilevel
the inspiratory trigger, can be set from 1 to 10, 1 is
S & ST
most sensitive, 10 is least sensitive.
Modes
This is a setting which controls the sensitivity of
Bilevel
the expiratory trigger, can be set from 1 to 10, 1 is
S & ST
most sensitive, 10 is least sensitive.
Modes
This is a setting which affects the backup pressure
Bilevel
cycle rate. A new breath will be triggered if the
ST & T
patient's breathing is slower than this setting.
Modes
This is a setting which affects the back up pressure
cycle rate. This is the ratio of the inhale to exhale
Bilevel
time and is expressed as 1:X. A new exhale will
ST & T
be triggered by the Bilevel if the patient's breathing
Modes
is slower than this setting, also linked to Backup
Rate. % inhale is shown in parenthesis.