INFORMATION MODE

Information mode allows you to view the Hour Meter and Serial Number.
ENTERING INFORMATION MODE
With the device plugged in and OFF:
- Press and hold all 3 buttons for 2 seconds until display shows Hour Meter.
Press Power button for 1 second and release for Serial Number.
Press Power button for 1 second and release for Firmware Version.
EXITING INFORMATION MODE
To Exit information Mode:
Press Power button for 2 seconds and release to exit Information Mode and turn off device.
NOTE– Information Mode times out after 2 minutes without user interaction.

ELECTROMAGNETIC COMPATIBILITY INFORMATION

If the device performance is lost or degraded due to electromagnetic disturbances, a visual and audible alarm indicates that the device is not meeting specification or a failure has been
detected.
WARNING
MR Unsafe
Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the iGo2 or MR medical
devices. The device and accessories have not been evaluated for safety in an MR environment.
Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security systems (metal
detectors) as it may cause unacceptable risk to the patient or damage to the iGo2. Some electromagnetic sources may not be apparent, if you notice any unexplained
changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and discontinue use. Contact your home care provider.
This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for
magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment
and the other equipment should be observed to verify that they are operating normally.
CABLES AND MAXIMUM LENGTHS OF CABLES
• DC power cord (cigarette lighter adapter) #DV6X-619 maximum length = 2.2 meter (7.2 ft)
EN - 16
Hr (Hour Meter)
Hr (Hour Meter) is displayed; it contains 5 digits shown across 3 segments. Display auto scrolls through
segments.
Fr (Firmware)
Fr (Firmware) is the next item. It contains 2 digits indicating the device's Firmware Version.
Sn (Serial Number)
Sn (Serial Number) is next item displayed; it contains 8 digits shown across 4 segments.
Display auto scrolls through segments.
SE-125-1

Publicité

Table des Matières
loading

Table des Matières