Hach DR 6000 Manuel D'utilisation De Base page 165

Masquer les pouces Voir aussi pour DR 6000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 229
Set up the wavelength scan
Press Wavelength Scan in the Main Menu. Press Options for
Parameter Setup.
Table 5 Options during wavelength scan
Option
Description
More
Further options are displayed
Folder icon
To store the scan data
From the displayed list of stored scans, a record is selected
for use as a reference scan/superimposed scan. This can be
highlighted or shown in the background in comparison with
the actual measured scan.
Reference Off/On
Note: This option is only available when there are
stored scans with the same wavelength range and
step.
λ
Enter the wavelength spectrum and scan interval
This functions as a stopwatch. It helps to make sure that the
steps of an analysis are correctly timed (e. g. reaction times,
Timer icon
wait times, etc., can be exactly specified). When the
specified time has elapsed, a sound is emitted. The use of
the timer has no influence on the reading program.
Enables the switching between the display of tabular scan
data (wavelength/absorbance) and the graphical
View
representation of the curve.
Table/Graph
Note: "View Table" is activated after the first
measurement.
Table 5 Options during wavelength scan (Continued)
Option
Description
To select Track or Peak/Valley. The selection for this menu
Cursor Mode
option determines to which points on the graph the cursor
moves.
To send data to a printer, computer or USB memory stick
Send Data
(Type A)
The integral gives the area and the derivative of the integral
Integral: On/Off
gives the original function
Scale: In the automatic scaling mode, the y-axis is
automatically adapted so that the total scan is shown.
Scale & Units
The manual Scaling mode allows sections of the scan to be
displayed.
Units: Choice of absorbance or transmittance.
Single Reading: One measurement result is shown after
tapping on Read.
Carousel 1 inch square: Optional measurement of carousel
Reading Mode
insert with up to 5 square cells.
Carousel 1 cm square: Optional measurement of carousel
insert with up to 7 square cells.
Recall
Recall saved measurement data, wave-length scans or time
measurement
courses, see
Save, recall, send and delete data, page
data
Instrument Setup
Basic data of the instrument, refer to
mode
delete data, page
Save, recall, send and
156.
English 165