4.
Remove the zero solution cell from the cell compartment. Insert the
sample cell into the cell compartment.
5.
Press Read. The result will be displayed.
Note: To define a sample dilution, press the Dilution key in the
toolbar.
6.
For data storage, refer to
Barcode Programs
A special barcode reader in cell compartment #1 automatically reads the
barcode on the 13 mm cuvette/vial as the cuvette/vial completes a single
rotation. The instrument uses the barcode identification to automatically
set the correct wavelength for the analysis and calculates the result
immediately with the help of stored factors.
In addition, readings are recorded at 10 different positions during a
rotation. A special outlier-elimination program is run and then the
average of the measured values is calculated. Cell and round cell faults
and contaminations are detected and therefore a very precise result is
determined.
English 160
section The data log, page
156.
Perform a barcode test
1.
Prepare the barcode test in accordance with the work instruction
and insert the cell into the cell compartment (1).
When a coded cell is inserted into the cell compartment (1)
(How to position the universal cell adapter, page
corresponding measurement program is automatically
activated in the main menu.
152), the