The following Safety Devices are installed:
- Safety valve set at 2.4 bar 0/+10%
- Electromagnetic lock to prevent the door from opening while the cycle is running
- Resistance over temperature thermostats
2.3
- The manufacturer is liable for the marketed product in accordance with current
regulations.
The manufacturer's liability will expire when operations are carried out on the device, or a
part of it, by unskilled personnel or using non-original spare parts.
- There should be no potential risk of explosion and/or fire in the room where the
autoclave is installed.
- The autoclave should be installed in a special well-ventilated room.
The incubator's connector (PICT.1 pos.18.), should be used only for the B-TEST.
The 230 V a.c. connection is indipendent by the MAINS, don't touch the connector with
wet hands and protect by the water.
SAFETY
Safety Marking
HAZARDOUS VOLTAGE
DISCONNECT THE POWER BEFORE REMOVING THE
LID
HIGH TEMPERATURE
EARTH CONNECTION
Safety Devices
Safety Notes