Press the MODE key (blue) J to enable the continuity test with buzzer/LED.
-
Bring the safety measuring leads into contact with the measuring points. If
the line resistance between the COM jack N and the jack O falls below
a value between 20 Ω and 200 Ω, the integrated buzzer of the BENNING
CM 10-PV sounds and the red LED 3 lights up.
see fig. 7:
Resistance measurement/ continuity and diode test/ capacitance
measurement
8.6 Capacitance measurement (switch position
Discharge capacitors fully before measurement! Never apply
voltage to the sockets for capacitance measurement as this may
cause irreparable damage to the unit. A damaged unit may repre-
sent an electrical hazard!
-
Use the rotary switch 6 to select the required function
CM 10-PV.
-
Determine the polarity of the capacitor and discharge it completely.
-
Connect the black safety measuring lead to the COM jack N of the BENNING
CM 10-PV.
-
Connect the red safety measuring lead to the jack + O of the BENNING
CM 10-PV.
Press the MODE key (blue) J twice to enable the capacity measurement
-
function.
-
Contact the discharged capacitor with the safety measuring leads observing
correct polarity. Read the measurement value on the digital display K of the
BENNING CM 10-PV.
see fig. 7:
Resistance measurement/ continuity and diode test/ capacitance
measurement
8.7 Diode test (switch position
-
Use the rotary switch 6 to select the required function
CM 10-PV.
-
Connect the black safety measuring lead to the COM jack N of the BENNING
CM 10-PV.
-
Connect the red safety measuring lead to the jack + O of the BENNING
CM 10-PV.
Press the MODE key (blue) J 3 times to enable the diode test.
-
-
Bring the safety measuring leads into contact with the measuring points
and read the measured value on the digital display K of the BENNING
CM 10-PV.
-
For a normal silicone diode located in flow direction, the flow voltage be-
tween 0.400 V and 0.800 V is displayed. If "000" appears in the display, there
may be a short circuit in the diode, „OL" indicates an interruption inside the
diode.
-
For a diode applied in reverse direction, „OL" is displayed. If the diode is
defective, „000" or other values are displayed.
see fig. 7:
Resistance measurement/ continuity and diode test/ capacitance
measurement
8.8 Temperature measurement (switch position )
-
Use the rotary switch 6 to select the required function
CM 10-PV.
-
Press the blue key J to switch over to °F or °C.
-
Connect the temperature sensor correctly to the COM jack COM N and jack
O ; observe polarity.
-
Place the contact point (end of the sensor lead) on the point to be mea sured.
Read the measurement value on the digital display K of the BENNING
CM 10-PV.
see fig. 8:
Temperature measurement
8.9 Current measurement with the flexible AC current transformer
BENNING CFlex 1 (switch position
-
Use the rotary switch 6 to select the required function
CM 10-PV.
-
Connect the black safety measuring lead to the COM jack N of the BENNING
CM 10-PV.
-
Connect the red safety measuring lead to the jack + O of the BENNING
CM 10-PV.
-
Select the 3,000 A (1 mV/A) measuring range at the AC current transformer
BENNING CFlex 1.
03/ 2020
)
BENNING CM 10–PV
)
)
on BENNING
on BENNING
on BENNING
on BENNING
37