3. Cycle the Resuscitator through several ventilations. The test lung should inflate during inspiration and
deflate during exhalation. Check for leakage at all joints and connections. Ensure that the Resusci-
tator refills promptly and properly and that all valves are operating correctly. If not, repeat the tests
above to find where the problem lies.
Specification & Performance
Storage Temperature: -40°C (-40°F) to 60°C (140°F)
Operating Temperature: -18°C (0°F) to 50°C (122°F)
Materials
Silicone rubber
silicone bag
duckbill valve
flapper valves
relief valve seal
infant mask
child mask
adult mask bladder
mask grommet
mask retainers
O-ring
Polyvinylchioride
oxygen reservoir bag
Stainless Steel
pressure relief valve spring
Connectors
patient port:
SILICONE BAG INLET
reservoir valve
intake valve port
Supplemental gas inlet
Polycarbonate
non-rebreathing valve housing
bag intake valve housing
reservoir valve housing
reservoir bag connector
adult mask shell
pressure relief valve housing
pressure relief valve stem
15 mm I.D. (F) / 22 mm O.D. (M)
23 mm I.D. (M)
25 mm I.D.
25 mm O.D.
6 mm O.D. (M)
ENGLISH