Conductivity; Basic Operation - Hach PAT700 DS Manuel D'utilisation De Base

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26
Figure 8 Flow control valve
1 Flow control valve

Basic operation

Select the run mode
1. Push << to show the sliding menu.
2. Push RUN MODE.
3. Select an operation.
Option
Online TOC

Conductivity

Standby
Offline
TOC manual sample
Self clean
4. Push ENTER.
Calibration
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current material safety data
sheets (MSDS) for safety protocols.
TOC calibration
Calibrate the instrument for TOC measurements with default settings or with custom settings. Use
default settings with packaged standards kits that use RFID tags on the bottles. Use custom settings
for less replicates or less bottles.
1. Make sure that the TOC standards are at room temperature.
2. Push << to show the sliding menu.
3. Push BOTTLE MODE, then RUN STANDARDS.
2 Water filter
Description
Monitor TOC continuously in process
Monitor conductivity continuously in process
Put the instrument in standby mode (valve open)
Put the instrument in offline mode (valve closed)
Manual TOC measurement for online sample
Configure and start a clean cycle
C A U T I O N
English 17

Publicité

Table des Matières
loading

Table des Matières