Télécharger Imprimer la page

Hach LANGE sensION MM374 Manuel D'utilisation page 3

Masquer les pouces Voir aussi pour LANGE sensION MM374:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 65
Table of contents
Specifications
on page 3
General information
on page 4
Installation
on page 5
User interface and navigation
on page 8
Startup
on page 9
Standard operation
on page 9
Specifications
Specifications are subject to change without notice.
Specification
Details
Dimensions
35 x 20 x 11 cm (13.78 x 7.87 x 4.33 in.)
Weight
1100 g (2.43 lb)
Meter enclosure
IP42
Power requirements
100–240 V, 0.4 A, 47-63 Hz
(external)
Meter protection class
Class II
Pollution degree
2
Installation category
Category II
Altitude requirements
Standard 2000 m (6562 ft) ASL (Above Sea Level)
Storage temperature
–15 to +65 °C (5 to +149 °F)
Operating temperature
0 to 40 °C (41 to 104 °F)
Operating humidity
< 80% (non-condensing)
Advanced operation
on page 12
Data logger
on page 13
Maintenance
on page 15
Troubleshooting
on page 17
Replacement parts and accessories
on page 18
Standard solutions
on page 19
Specification
Details
Resolution
pH: 0.1/0.01/0.001, ORP: 0.1/1 mV, ISE:
programmable, temperature: 0.1 ºC (0.18 ºF), EC:
variable, resistivity: variable, NaCl: variable, TDS:
variable
Measuring error (±
pH: ≤ 0.005, ORP: ≤ 0.2 mV, temperature: ≤ 0.2 ºC (≤
1 digit)
0.36 ºF), EC: ≤ 0.5%, resistivity: ≤ 0.5 %, NaCl: ≤
0.5 %, TDS: ≤ 0.5%
Reproducibility (±
pH: ± 0.001, ORP: ± 0.1 mV, temperature: ± 0.1 ºC (±
1 digit)
0.18 ºF), EC: ± 0.1%, resistivity: ± 0.1%, NaCl ± 0.1%,
TDS ± 0.1%
Data storage
330 results and last 9 calibrations
Connections
2 Combined or indicator probe: BNC connector (Imp.
12
>10
Ω); 2 Reference electrodes: banana connector;
A.T.C. type Pt 1000: banana or telephonic; 2 magnetic
stirrers: RCA connector
Conductivity probe with built-in Pt1000 sensor (or NTC
10 kΩ probe): telephonic connector
RS232C for printer or PC: telephonic connector;
external PC keyboard: mini DIN connector
Temperature correction Channel 1 pH: Pt 1000 temperature probe (A.T.C.),
NTC 10 kΩ probe, manual, Channel 2 temperature,
isopotential pH programmable, standard value 7.00,
CE: Pt 1000 temperature probe (A.T.C.), linear
function, TC=0.00 to 9.99%/Temperature, TRef: 20°C
(68 °F) or 25°C (77 °F), non-linear function for natural
waters (UNE EN 2788) Channel 2 pH: Pt
1000 temperature probe (A.T.C.), manual, NTC 10KΩ
probe, isopotential pH programmable, standard value
7.00
Measurement display
Continuous measurement, by stability, by time
lock
Display
Liquid crystal, backlit, 128 x 64 dots
Keyboard
PET with protective treatment
Certification
CE
English 3

Publicité

loading