Ormed ARTROMOT-SP3 Mode D'emploi page 71

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 76
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
®
ARTROMOT
-SP3 is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the ARTROMOT
is used in such an environment.
Immunity test
IEC 60601
test level
Conducted RF to
3 V
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF to
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radio, AM and FM radio broadcast and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due
to fixed RF transmitters, an electromagnetic site survey should be considered. If the mea-
sured field strength in the location in which the ARTROMOT
the applicable RF compliance level above, the ARTROMOT
ved to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the ARTROMOT
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance level
3 V
rms
rms
3 V/m
71
®
-SP3 device
Electromagnetic environ-
ment-guidance
Recommended separation di-
stance:
r
1.2
P
ö
d =
r
1.2
P
ö
d =
80 MHz to
800 MHz
r
2.3
P
ö
d =
800 MHz to
2.5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey
is less than the compliance
level in each frequency range
Interference may occur in the
vicinity of equipment marked
with the following symbol
®
-SP3 device is used exceeds
®
-SP3 device should be obser-
®
-SP3 device.
d
a
,
b
.

Publicité

Table des Matières
loading

Produits Connexes pour Ormed ARTROMOT-SP3

Table des Matières