5. Setting the treatment values

Warning
Patient hazard –
Before therapy, a test run consisting
of several exercise cycles must be
completed without the patient. Then
repeat the test run with the patient
and check that the movement does
not cause any pain.
Note!
See also 2.2 and 2.3 as well as page
221!
5.1 General information
on programming
ARTROMOT
1. You activate the programming mode
by briefly pressing the MENU key on
the programming unit.
2. The treatment parameters and func-
tions are allocated to four (Standard
model) or six (Comfort model) pro-
gramming levels (four per level).
To be able to program a parameter
you will have to access the corre-
sponding programming level. This is
also done with the MENU key. With
each key press you advance one
level. The code M1, M2, etc. that ap-
pears in the middle of the display in-
dicates the programming level.
3. You activate the treatment parame-
ters and functions with the four pa-
rameter keys below the display. The
symbols above the four parameter
keys indicate the assigned parame-
ters and functions.
®
-SP3
54
This is what happens when you press
one of the parameter keys to select a
parameter:
- The corresponding symbol appears
on the display in a larger format.
- The set value is displayed.
- The symbol above the parameter
key appears in reverse video.
4. With the + / - keys (plus/minus) you
change the displayed value. When
you press and hold the key, the value
will change at a higher rate.
Some of the (special) functions can
only be enabled and disabled. This is
done by pressing the corresponding
parameter key or with the + / - keys.
Active parameters are identified with
a check mark in the circle next to the
symbol.
5. Having programmed all parameters,
press the STOP key to save the val-
ues.
6. Then press the START key:
ARTROMOT
®
-SP3 will check the set
values, it moves to the middle posi-
tion and stops.
7. Press the START key again to start
therapy.
In the synchronized mode, the car-
riage will then move from the middle
position to the maximum eversion
value. Afterwards, the carriage will si-
multaneously move to the maximum
inversion value and to the maximum
plantar flexion value, and then simul-
taneously to the maximum dorsal ex-
tension value and the value halfway
between inversion and eversion. After
reaching this position, the cycle
restarts: maximum eversion at the
same time as halfway between dorsal
extension / plantar flexion, etc.
After activation of the START key in
the non-synchronized mode the mo-
tors will perform random movements,
each motor reversing after reaching
the maximum values.