Lift position key (4&6) (Fig. 23a and Fig. 23b)
4
Assembling the stand pillar RV8.1 (Accessories)
No stand is required if the condenser has been properly fitted
and secured by tightening the condenser union nut on the rotary
drive.
The purpose of the stand is solely to prevent the condenser
twisting.
WARNING
Caution: If the RV8.1 stand is fitted incorrectly, the strong forces
at the Velcro strap can cause stresses within the glass, which in
turn can cause damage to the glass condenser.
After the condenser has been fitted, the RV8.1 stand is attached
to the rotary drive. Make sure that the condenser is installed
parallel to the lift body.
- Guide the stand rod through the hole on the underside of the
rotary drive,
- Screw the knurled screw into the tapped hole but do not tighten
it,
- Move the stand until the upper rubber sleeve rests against the
glass condenser,
- Tighten the knurled screw finger-tight,
- Secure the condenser to the stand using the pre-fitted Velcro
strap.
To adjust the lift position, grasp the handle and press the left or
right lock button (4).
The hand must rest on the sensor (6).
If the sensor (6) is touched for longer than 60 seconds (hand
contact), the locking of the lift position is activated permanently.
6
The lift position cannot be adjusted, despite pressing the left or
right lock button (4).
To reactivate the sensor (6), remove the hand. Then the lift-
position can be adjusted as described.
Note: Hand detection sensor (6), only to be touched by hand,
do not press!
Fig.23a
The lift lock is unlocked and you can move the lift up and down.
Release the lock button or remove the hand from the lift, the lift
is locked in the adjusted position.
When locking and unlocking, a clear "click" sound is heard.
Fig.23b
Condenser
Stand pillar RV 8.1
Fig.24
Rotary drive
30