Télécharger Imprimer la page

Inserting And Removing The Rotor; Loading The Rotor - IKA mini G Mode D'emploi

Masquer les pouces Voir aussi pour mini G:

Publicité

Opening the cover
To open the cover, press the "Release cover" but-
ton (item C).
Closing the cover
Press the cover down manually until it clicks into
place.

Inserting and removing the rotor

Slacken or remove the rotor securing nut (item
A), then place the rotor on the motor shaft. It
must slide on until it reaches a clearly felt stop.
Now secure the rotor by tightening the rotor
nut (finger-tight clockwise). Restrain the rotor by
hand while the nut is being tightened.
Before starting the centrifuge, always check that
the rotor is securely fastened.

Loading the rotor

see Fig. 8.
in the device.
the maximum amounts stated by the manufac-
turer.
The cover opens with a damped movement.
The cover should close easily.
To undo the rotor, restrain the rotor by hand and
then unscrew the rotor nut anticlockwise. If the
rotor nut is too tight, a tool (such as a screwdri-
ver) can be used to undo it.
as to keep the difference in weight between
centrifuging vessels as small as possible. Tare
the centrifuging vessels using scales.
Fig. 8
16

Publicité

loading