SAFETY RULES

Observe all the instructions set forth in this manual.
The dental lamp is used to illuminate the operating area and should
be used in a dental office by medical personnel only.
The device should be cleaned before use (see para. 'Cleaning').
The lamp packing is designed to adequately protect it from
penetration of external agents.
The device should be installed by qualified personnel only.
The dental lamp should be installed on a specific power control
device such as a dental unit or with an electrical system meeting IEC
standard 364-1 and 'national rules for the installation of electrical
systems in rooms assigned to medical use'.
The device should be installed with a double-pole switch for the net
separation (Conforms to publication IEC 328)
Make sure the power supply shown on the data plate is the same as
that of the mains.
Do not perform any maintenance on the lamp when the power is on.
Disconnect the power cord from the mains before repairing.
Do not insert into the slots in the lamp head objects or tools which
might touch parts under voltage.
Inadequate performance of the device does not endanger the patient
in any way
Separate materials according to their typology (ex. ferrous, rubber, plastic
etc.). For the material scrapping and disposal follow the standards in force
in your country or apply to specialized and authorized companies. Do not
leave small parts of the device unattended or within the range of children
because dangerous.
Arm designed load 5 Kg