Parameter
Calibration date
Calibration initials
Calibration period
(days)
Measurement location
ID
1)
Typical values for the CFT/20kN force transducer
CAUTION
Please note that replacing the sensor or changing the system gain
requires adaptation of the TEDS data. Downstream systems would
otherwise use incorrect sensor settings.
Therefore, when replacing the sensor or changing the system gain, please
adapt the TEDS data.
For more extensive information about TEDS, look in the TEDS Operating
Manuals on the Internet at www.hbm.com/TEDS
A2781−2.1 en/de/fr
1)
Value
Unit
27-June-2007
HBM
0
days
0
Change
requires
rights to
level:
CAL
Date of the last calibration or
creation of the test certificate (if
no calibration carried out), or of
the storage of the TEDS data (if
only nominal (rated) values from
the data sheet were used).
Format: day-month-year.
Abbreviations for the months: Jan,
Feb, Mar, Apr, May, Jun, Jul, Aug,
Sep, Oct, Nov, Dec.
CAL
Initials of the calibrator or
calibration laboratory concerned.
CAL
Time before recalibration,
calculated from the date specified
under Calibration Date.
USR
Identification number for the
measuring point. Can be assigned
according to the application.
Possible values: a number from 0
to 2047. If that is not enough, the
HBM Channel Comment template
can also be used for this purpose.
23
Explanation