Factory Settings; Alarm - Henry Schein 900-8275 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ENGLISH
5.4.5 Transfer Set...
Pick and enter [Transfer Set] interface from the [System Setup] interface, refer to Fig.5-12. In this interface
there are two transmission modes for selection: Bluetooth and USB. But the Bluetooth transmission is
reserved.

5.4.6 Factory Settings

Pick and enter [Factory Settings] interface from the [System Setup] interface, refer to Fig.5-13.
Press the MENU button to highlight the item, and then press the UP or DOWN navigation button to change it
between YES and NO, at last press the MENU button again to confirm your setting.
If you select YES the default settings will resume as follows:
Alarm: ON
Beep: ON
Display Mode: WAVE
Brightness: 60%
Transfer Set...: USB
Sleep Mode Sleep status: OFF; Sleep time: 60 second.
SpO
alarm upper limit: 100%; lower limit: 95%.
2
PR alarm upper limit: 100bpm; lower limit: 60bpm.
ALARM PRIORITY:
There are three-level priorities for selection.
High priority: Highest level alarm indicates the patient is in a very dangerous situation.
Medium priority: indicates that warning-attention should be paid.
Low priority: indicates the technical alarm caused by the device itself.
Alarms of the oximeter include technical and physiological alarms. All the three priorities divided by built-in
module and cannot be changed by user.
Assignment of priority:
Parameter
Fig.5-11
Fig.5-12
Fig.5-13

6. Alarm

High
Medium
SpO
2
PR
11
Low

Publicité

Table des Matières
loading

Table des Matières