Other Settings; Alarm Setup; History - Henry Schein 900-8275 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ENGLISH
methylene blue
o
Indigo carmine
o
4.3.3 Extremely high illumination could affect the SpO
shield the sensor. Pulse sensor should obviate the light source, e.g. radial lamp or infrared lamp.
4.3.4 Other factors
High-frequency electrosurgical interference from external devices, including defibrillators.
o
Placement of a probe on an extremity that currently has installed a blood pressure cuff, arterial catheter,
o
or intravascular line;
The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia;
o
An arterial occlusion proximal to the probe.
o
Warnings!
Use only SpO
probes provided by manufacturer. Other SpO
o
2
Do not use the SpO
o
Excessive patient movement may cause inaccurate measurements.
o
Tissue damage can be caused by incorrect operation or misusing probe; for example, by wrapping the
o
probe too tight. Inspect the probe site to ensure the skin's integrity and the adhesion position of the probe
is correct. More frequent inspection should be taken if necessary.
Loss of pulse signal can occur in any of the following situations:
o
The probe is too tight;
o
There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or
o
sunlight; Pulse sensor should obviate the light source, e.g. radial lamp or infrared lamp.
A blood pressure cuff is inflated on the same extremity as the one to which the SpO2 sensor is
o
attached.

5.1 Alarm Setup

Pick and enter the [Alarm Setup] interface from the main menu. And then press the Navigation buttons to
select different items to set. Press the MENU button to set and then adjust the item to your needs. Press the
MENU button again to confirm the settings.
Beep: By setting the "Beep" ON or OFF, you can set the pulse sound on or off.

5.2 History

Pick and enter the [History] interface from the main menu. Its sub-items are located at this interface, refer to
Fig.5-2.
5.2.1 Data Browse
Pick and enter the "Data Browse" interface as shown in Fig.5-3. By pressing the LEFT/ DOWN button, the
user can go to the earlier item page; By press the RIGHT/ UP button, the user can go to the later item page.
The pulse oximeter can record the alarm parameter marked with red color. Press the MENU button, it will
return to the previous interface.
probe with exposed optical components.
2

5. Other Settings

Fig.5-1
Fig.5-2
measurement. Use a semi-translucent or opaque cover to
2
probes may cause improper performance.
2
8

Publicité

Table des Matières
loading

Table des Matières