Troubleshooting
Causes and Corrective Measures
Error
Measuring tool cannot be
switched on
Measuring tool switched on
but does not react
Display indication:
"S l i p p i n g W h e e l"
Display indication:
"S p e e d i n g"
"Temperature over range"
"Temperature under range"
"Strong radio signal detected"
Subject to change without notice.
Bosch Power Tools
Cause
Batteries empty
Batteries incorrectly in-
serted (wrong polarity)
Measuring tool too warm or
too cold
Wheel losing contact with
the surface
Measuring tool has been
moved to quickly
English | 29
Corrective Measure
Replace batteries
Check if the batteries are inserted
correctly
Take out batteries and reinsert again
Wait until operating temperature
range is reached
Press the start button 11 and take care
that the two bottom wheels have con-
tact with the wall while moving the
measuring tool; in case of uneven walls,
position a thin piece of cardboard be-
tween the wheels and the wall
Press the start button 11 and move
measuring tool slowly over the wall
Wait until operating temperature
range is reached
Wait until operating temperature
range is reached
Measuring tool switches off automati-
cally. If possible, eliminate the interfer-
ing radio waves, e.g. WLAN, UMTS,
radar, transmitter masts or micro-
waves, then switch the measuring tool
on again.