Operation
Send data to the computer
Data that has been downloaded to the USB+Power module can be
transferred and stored to a computer. The data will be formatted as an
XML file.
1. Push the DATA MANAGEMENT key.
2. Select Send Data Log.
Option
Description
Send Data All data is sent to the USB+Power module
Filter Data Data is filtered by:
Data Type—All Logs, Reading Log, Calibration Log and Verify
Cal Log
Time Interval—All Logs, Last Reading, Current Day, Current
Week and Current Month
Sample ID—Off or Sample ID
Operator ID—Off or Operator ID
a. Select Send Data to send all data to the USB+Power module.
b. Select Filter Data. Make a choice of the data to transfer and push
Exit. Select Send Data to send the selected data to the USB+Power
module.
A progress bar will show the transfer status and shows "Transfer
Complete" when the data is successfully sent to the USB+Power
module.
3. Remove the USB+Power module from the meter.
Note: Do not connect the USB+Power module to the computer when the module
is still connected to the meter.
4. Plug the USB cable, type A, to the USB+Power module and connect
the USB+Power module to the PC.
Note: The USB+Power module works like a USB storage device. Navigate
through Windows Explorer to find the USB+power module.
5. Open the folder "datalog". Find the data file.
8 English
The file name will be formatted as: Year_Month_Day_Hour_Minute.
xml.
6. Save the data file to a location on the computer.
7. Open a spreadsheet program to view the data.
Print stored data
All data can be sent automatically to a printer. The PD-2 Citizen Printer is
compatible with the meter and is FCC Part 15B, Class B compliant. Other
printers may not be compatible.

Printer setup

Refer to the documentation supplied with the printer to select the USB
interface.
Print data
1. Turn off the meter. Install the USB+Power module to the meter and to
AC power.
2. Connect the printer to the USB+Power module with the USB cable
supplied with the printer and the special USB cable adapter supplied
with the module.
3. Connect the printer to power and turn the printer on.
4. Push the ON/OFF key to turn the meter on. The meter automatically
detects when the printer is installed.
Note: After a reading is taken, the data is automatically transferred to the printer
when the Auto-print data is selected.
5. Push the DATA MANAGEMENT key and select Send Data Log. Select
one of the following options:
Option
Description
Send Data Data is sent directly to the printer.