Hach 8000A Manuel D'utilisation page 11

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Cumulative Limit value is 9999999. This field is active only if Counts
Mode is selected.
10. Enter a value for the Sample Time between 00 H, 00 M, 00 S and
02 H, 59 M, 59 S. This value is the length of each sample run. This
field is active only if Time Mode is selected.
If Fed-Std-209E is selected, Sample Time is calculated from the
sample volume specified in the 209E SETUP function, assuming a
flow rate of 1.0 cubic feet per minute. The sample time cannot be
changed directly.
11. Enter a value for the Sample Volume between 0.1 and 9999 mL. This
is the quantity of fluid that will be sampled in each analysis. This
value is preset to 5 mL if USP24<788>, JP<13> or EP<99> is
selected and a glassware check is being run. For MANUAL and
TIME modes, set the sample volume to match the sample volume
delivered during the run.
12. Enter a value for the Pooled Volume between 20 and 9999 mL. This
value is the quantity of fluid removed from the Pooled Containers and
placed in a common container from which the sample is drawn.
Pooled Volume will vary depending on which standard is used. The
Pooled Volume field is available only when USP24<788>, JP<13> or
EP<99> is selected.
13. Enter a value for the Containers Pooled between 10 and 9999. This
is the number of containers the Pooled Volume is drawn from.
The value will vary depending on which standard is used. The
Containers Pooled field is available only when USP24<788>, JP<13>
or EP<99> is selected.
14. Select the Test Type, i.e., Glassware test or Sample test. The
glassware test makes sure that the environment is suitable for the
analysis, the glassware is clean, and the water used for analysis is
free of particles. Sample Test refers to the testing of the Sample
Volume.
The Test Type field is available only if USP24<788>, JP<13> or
EP<99> is selected.
Set passwords and access levels
The counter has three levels of operator access: High, Medium and Low.
Operators with High level privileges can access all the functions
available to operators with Medium and Low access. Low level access
does not require a password.
On start up, an operator must enter a password when prompted by the
instrument. If the operator pushes only the ENTER key, or enters an
invalid password, the operator has only low level access.
Make sure that the date and time for the counter are correct before
access levels and passwords are set.
1. In the Password functions menu, select an option.
Option
Description
SET LEVEL
Sets the access level lower than the current level. This is the
only password button enabled for medium level operators.
SET
Sets the passwords for high and medium level access. The
PASWD
password must be typed in once, then typed in again for
verification. Passwords are alphanumeric entries of up to
16 characters.
SET
Only available to high level users. Controls the functions
ACCES
available to medium and low level operators. The medium
level operator has access equal to or greater than the low
level operator access. If a high level operator grants a low
level operator access to a function, the medium level operator
access is automatically enabled.
2. Update or change the settings for the selected function.
Set the clock
Set the instrument clock as soon as possible after installation. Push
ENTER to accept the data in a field segment and move to the next one.
1. In the Miscellaneous functions menu, push SET CLOCK.
2. Enter the date and time data in the format MM/DD/YY HH:MM:SS.
English 11

Publicité

Table des Matières
loading

Table des Matières