WARNING!
– Cylinders that have been refinished and/or exposed
to elevated temperatures MUST be hydrostatically
tested before filling. Cylinder temperature should not
exceed 52° C (125° F). Aluminum cylinders subject to
the action of fire or heated temperatures in excess of
177° C (350° F) MUST be withdrawn from service
and condemned by trained personnel. DO NOT
use caustic paint strippers. Valve and safety relief
devices MUST be removed and replaced by trained
and authorized personnel. DO NOT alter or change
valve assembly.
– All cylinders MUST be inspected before attempting to
fill; otherwise, injury or damage may occur.
– NEVER smoke in an area where oxygen is being
administered.
– NEVER use near any type of flame or
flammable/explosive substances, vapours or
atmosphere.

3.3 The Hydrostatic Testing Date

Aluminum cylinders MUST undergo testing every five years.
WARNING!
– DO NOT fill cylinders that have not been tested
in the past five years. Otherwise, serious personal
injury may result. Contact your dealer/Home Care
equipment provider for replacement.
1176148-A

3.4 External Examination

1.
Examine the outside of the cylinder for the following conditions,
and replace the cylinder if they exist:
Dents or dings
Arc burns
Oil or grease
Any other signs of damage that might cause a cylinder to be
unacceptable or unsafe for use.
2.
Examine the cylinder for evidence of fire or thermal damage.
evidence includes charring or blistering of the paint, or other
protective coating or heat sensitive indicator. If fire or thermal
damage is found, replace the cylinder.
3.
Inspect the cylinder fillport for the following:
Debris, oil or grease
Noticeable signs of damage
Signs of corrosion inside the valve
Signs of excessive heat or fire damage
Signs of pressure leaks
DO NOT use the cylinder if any of these items are
found. Contact your dealer/Home Care equipment
provider for replacement.