OPERATING INSTRUCTIONS OF
VII - Samsoft 175 MAINTENANCE
Disconnect the power supply cord and remove the
battery pack before any maintenance operation.

1) Cleaning

(every month or when necessary, and to be performed
by a competent staff )
Clean the patient lift with a soft cloth dampened with
soapy solution, then rinse, and carefully dry.
Do not use aggressive detergent; do not soak.

2) Disinfecting the patient lift

(when necessary, and to be performed by a competent
staff )
Disinfecting is performed with detergents and disinfecting
solutions sprayed on the external chair areas.
It is a foaming solution that does not require rinsing, but
only drying. Our internal services use ANIOS solutions.

3) Daily maintenance

Charge the battery pack every night.

4) Routine maintenance

(every month or when necessary, and to be performed
by a competent staff )
The routine maintenance consists in checking the system
functionality:
With the patient lift positioned on a fl at fl oor, check its
4 wheels are touching the fl oor.
When the brakes are engaged, the patient lift should
not be moved (fl at and adhered fl oor).
Operate the patient lift with nobody in, check the arm
and the side rails move steadily with no noise.
Control the handles located on the mast and check
there is no play between the arm and the mast, and
between the mast and the side rails.
Check the plastic protective housings and casings are
not broken or cracked, and all parts are present.
Check the condition of the connecting cables: power
supply cord, remote control and cylinder; if a cable shows
a slight burn, wear, or impact sign, or if the wires are
visible, immediately stop using the patient lift, and have
the cables replaced by the engineering department.
The slings should be checked at least once a month,
or more regularly in case of intensive use, ensuring the
seams are not undone, the upholstery and sling weft is
not visible, and there is no torn part.
Never use a defective sling.
The worn or malfunctioning parts should be
replaced with original parts manufactured by
Drive DeVilbiss Healthcare France.
Have some tries with a load before commissioning the
patient lift again.
22
Samsoft 175

5) Preventive maintenance

Every year or when necessary, and to be performed by
a competent staff , according to paragraphs 5.1, 5.2, and
5.3 below.
It is necessary to control the device every year.
The battery pack should be changed about every
18 months depending on how it is used.
The crossbar suspension axle should be imperatively
checked yearly on each maintenance task. Yearly
maintenance should be performed by skilled technical
staff experienced with electromechanical device
maintenance. The maintenance tasks do not require
particular tools; adequate tools should be used according
to the best industry practices. Mechanical fastening,
plays, and adjustments should be realised according to
the best industry practices
5.1) Mechanical control
Disassembly and systematic lubrication of the
crossbar axle. In case of any wear sign, the axle should
IMPERATIVELY be replaced.
Dismantling, cleaning, and lubrication of the front and
rear casters.
Control and lubrication after dismantling the mast and
side rail pivots.
Check the welding. Visual inspection.
Check the alignments. Visual inspection.
Control and lubrication of the cylinder fastenings.
Check the tightening of all the parts.
Dismantling, cleaning, and lubrication of the mast
hinge. Check the fastening, caster play, and brake
performance. The mechanical parts that have cracks,
wear signs or geometrical defects should be replaced
with original components delivered by Drive DeVilbiss
Healthcare France.

5.2) Electrical control

Check the remote control.
Check the charger.
Check the remote control box and the cylinder.
Check the lifting/lowering speed.
Check the emergency stop and lowering.
Condition of the electric cables (change the cables in
case of doubt).
1) Check the axle is not distorted.
2) Check the collar is not worn.
EN