Identification Label; I - Specifi Cations; Samsoft 175 - Drive DeVilbiss Healthcare Samsoft 175 Guide D'utilisation

Masquer les pouces Voir aussi pour Samsoft 175:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

OPERATING INSTRUCTIONS OF
Check the red emergency lowering ring is in
the up position.
Check the elastic ring is correctly engaged
in the groove of the axle.
Check the spring pin is correctly oriented on
either side of the lug and correctly closed.
Check the beta pin in the lower part of the mast
is correctly positioned.
Check the wing knob is correctly screwed in.

IDENTIFICATION LABEL

Manufacturing date (week/year)
I - SPECIFICATIONS

1. Samsoft 175

Dimensions (folded)
Dimensions (unfolded)
Lifting capacity
Inner feet spacing
Turning diameter
Base height
Lifting time
Patient lift weight
Lifetime
Front wheels
Rear wheels
Structure
Samsoft 175
Lxlxh 130 x63x40 cm
Lxlxh 133x63x134 cm
175 kg
54 cm min to 83 cm max.
144 cm
9 cm
38 s (empty)
38,4 kg
10000 cycles at 175 kg
Ø 7,5 x 6 cm without brake
Ø 7,5 x 6 cm with brake
Steel
Coating
Electric cylinder (lifting)
Mechanical
cylinder
rod
disengagement (red ring)
Battery power supply
External battery charger
Battery charge indicator
Buttons
Cycle number as for patient lift
autonomy (lifting/lowering) with
an average weight of 100  Kg,
with new charged batteries.
EC marking date
Complies with ISO 10535 standard
Important: the patient lift should be used with
slings (safety members) provided by Drive
DeVilbiss Healthcare France exclusively.
83 cm
63 cm
54 cm
73 cm
Dimensions +/- 1 cm
Epoxy RAL 7005 - 7004 - 8000N
24 V continuous 8000N
2 x 4.5 Ah, 12 V (included in the
battery pack)
Par LED
Emergency stop and emergency
lowering
80 cycles
2014
Driven operating force :
with one fi nger : ............. < 5 N
with one hand : ..........< 105 N
with one foot : ...........< 300 N
54 cm
Pointing to the front
EN
9 cm
15

Publicité

Table des Matières
loading