OPERATING INSTRUCTIONS OF

2. Lifting assembly

Autonomous box : (control box + battery pack)
Dimensions
Weight
Protection
Output voltage
Output current
Output number
Battery voltage
Battery capacity
Autonomy
Charging time
Emergency stop push-button/
Emergency lowering button
Battery indicator
Protection index
Actuator of cylinder :
Power supply
Max center to center distance
Max. stroke
Max. thrust
Lifting time when empty
Lifting time when a patient is in 54 s
No-load speed
On-load speed
Weight
Protection index
Mechanical decoupling /
Safety nut
Sound level
External battery charger
Power supply
Output
Protection index
CERTIFICATION
Samsoft 175 complies with ISO 10535 standard (March
2007).
16
Samsoft 175
L=368 mm l=120 mm h=107,6 mm
4,8 kg
Current limiter
24 V continuous
8,5 A maxi
1
2 x 12-V batteries
4,5 Ah
Around 80 cycles
8 h
On the control box
Charged/discharged
IP 54
24 V continuous
895 mm
300 mm
8000 N
38 s
26 mm/s
19 mm/s
3,5 kg
IP X4
≤ 60 dB
:
230 VAC 50-60 Hz
24 V continu
IP 20
II - INTENDED USE
This transfer device is intended to transfer mobility-
impaired patients; it can be used with a wheelchair, a
shower seat, a bed, or a toilet seat.
Not suitable for children under 12  year-old, for
patients with mental disorders, and for patients
whose morphology is incompatible with this kind
of transfer device
III - PRINCIPLES OF ASSEMBLY

1) General terminology

4
5
6
8
7
9
11
1 - Mast
2 - Base
3 - Cross-bar
4 - Boom
5 - Lifting cylinder
6 - Remote control
7 - Control box
EN
3
10
2
8 - Battery pack
9 - Spacing pedal of
the side rails
10 - Side rail
11 - Locking system of the
side rails in the open
position (optional)