PC Diagnostic Tool
Start diagnostic tool
10.1
Test
Customer Diagnostics
Technician Diagnostics
Sirona Windows Diagnostics
Tip: If one of the following steps does not work, further steps can be
found in the Service Manual.
ü The PC is switched off.
1. Switch on the PC.
2. Repeatedly press the "F11" or "F8" button (in V2.2.1 - V2.2.3) until
the PC switches to the boot menu.
3. Insert the diagnostics tool CD into the drive.
4. Select the "P3: TSSTcorp CDDVDW ..." menu option.
5. Confirm your selection with the "Return" key on the keyboard.
Ä The diagnostics tool starts and a selection menu appears.
Tip: After one minute, if none of the arrow keys have been
pressed, the "Customer Diagnostics" test starts
automatically.
6. Select the test using the arrow keys.
7. Confirm your selection with the "Return" key on the keyboard.
Ä The test starts.
Test procedure
10.2
10.2.1
Customer Diagnostics
Suitable for
Suitable for the user, in order to
check the PC components.
Suitable for the technician, in order to
check the PC components.
Windows Stress Test, in order to test